Technique | Product | Hydrogen Carrier Gas | Nitrogen Carrier Gas |
---|---|---|---|
GC | 8890 GC | Yes (all inlets) | Yes (all inlets) |
Intuvo 9000 GC | Yes (all inlets) | Yes (all inlets) | |
8860 GC | Yes (all inlets) | Yes (all inlets) | |
8850 GC | Yes (all inlets) | Yes (all inlets) | |
990 Micro GC | Yes | Yes | |
490 Micro GC | Yes | Yes | |
7890 Series GC | Yes (all inlets) | Yes (all inlets) | |
7820 Series GC | Yes (all inlets) | Yes (all inlets) | |
6890 Series GC | Yes (all inlets) | Yes (all inlets) | |
6850 GC | Yes (all inlets) | Yes (all inlets) | |
GC Detectors | FID | Yes (capillary columns) | Yes (capillary and packed columns) |
TCD | Yes (capillary and packed columns) | Yes (capillary and packed columns)* | |
ECD | Yes (capillary columns) | Yes (capillary and packed columns) | |
NPD | No | Yes (capillary and packed columns) | |
FPD | Yes (capillary and packed columns) | Yes (capillary and packed columns) | |
SCD | Yes (capillary and packed columns) | Yes (capillary and packed columns) | |
NCD | Yes (capillary and packed columns) | No | |
GC/MS | 5977 Series single quad | Yes | Not recommended |
7000 Series triple quad | Yes | Not recommended | |
7010 Series triple quad | Yes | Not recommended | |
7250 Q-TOF | No | Not recommended | |
7200 Series Q-TOF | No | Not recommended | |
5975 Series single quad | Yes | Not recommended | |
5973 Series single quad | Yes | Not recommended | |
Sample introduction | |||
Headspace | 8697 Series | Yes | Yes |
7697A Series | Yes | Yes | |
G1888 | No | Yes | |
7694 | No | Yes | |
Thermal Desorption | TD-xr | Yes (multi-gas version) | Yes (all versions) |
Purge and Trap | Lumin | Yes (must use inert gas for purge) | Yes |
AQUATek | Yes (must use inert gas for purge) | Yes | |
AtomX | Yes (must use inert gas for purge) | Yes | |
* Typically only used when measuring helium or hydrogen in a sample. Information in this table is current as of April 24, 2024. |
Pros | Cons | |
---|---|---|
Helium Carrier Gas | Always the first choice, if available | Frequent shortages and delivery interruptions |
Excellent chromatographic and MS performance | Inconsistent or high costs | |
All reference spectra in libraries are obtained with helium | Sometimes difficult to find in chromatographic grades | |
Hydrogen Carrier Gas | Best alternative to helium | Requires attention to safety |
Better chromatographic resolution and speed compared to helium | Requires attention to reactivity | |
Nitrogen Carrier Gas | Inexpensive | Requires longer run times to achieve resolution |
Widely available in chromatographic grades | Not recommended for use with MSD | |
Safer than hydrogen |
Method translator is built into OpenLab CDS software, or it can be downloaded as a standalone application.